Teledyne ML – T101 Operation Manual
Calibration Procedures
183
It is recommended that, if dynamic calibration is enabled, the analyzer’s test functions,
slope and offset values be checked frequently to assure high quality and accurate data
from the instrument.
6.9.1.
AUTOCAL OF INSTRUMENTS IN INDEPENDENT OR AUTO
REPORTING RANGE MODES
If the analyzer is being operated in
IND
or
AUTO
Range mode, then the High and Low
ranges must be specified as part of the Auto Cal set up. This parameter appears at the end
of the programming sequences after the
CALIBRATE: ON/OFF
parameter is set. For
example:
SETUP X.X
RANGE TO CAL: LOW
<SET
EDIT
EXIT
EXIT
returns
to the
SETUP
Menu
SETUP X.X
CALIBRATE: ON
<SET
SET>
EDIT
EXIT
SETUP X.X
SEQ 2) ZERO–SPAN, 2:00:30
PREV NEXT
MODE
SET
EXIT
Sequence
Delta Time
MODE Delta Days
Follow standard AutoCal programming process to this
point, then …
SETUP X.X
RANGE TO CAL: LOW
LOW
HIGH
ENT
R EXIT
SETUP X.X
RANGE TO CAL: HIGH
LOW HIGH
ENTR
EXIT
SETUP X.X
RANGE TO CAL: HIGH
<SET
EDIT
EXIT
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...