Operating Instructions
Teledyne ML – T101 Operation Manual
98
4.6.5.
ELECTRICAL TEST
The electrical test function creates a current, which substitutes the PMT signal, and feeds
it into the preamplifier board. This signal is generated by circuitry on the pre-amplifier
board itself and tests the filtering and amplification functions of that assembly along with
the A/D converter on the motherboard. It does not test the PMT itself. The electrical test
should produce a PMT signal of about 2000 ±1000 mV.
To activate the electrical test press:
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL
SETUP
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
DIAG
SIGNAL I / O
NEXT
ENTR
EXIT
DIAG ELEC
RANGE = 500.000 PPB O2=X.XXX
<TST TST>
EXIT
Press
NEXT
until…
DIAG
ELECTRICAL TEST
PREV NEXT
ENTR
EXIT
Press
TST
until…
DIAG ELEC
PMT = 1732 MV
SO2=X.XXX
<TST TST>
EXIT
While the electrical test is
activated,
PMT
should equal:
2000 mV ± 1000 mV
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...