Teledyne ML – T101 Operation Manual
Calibration Procedures
171
Step Two:
Set the expected H
2
S
span gas
value:
This sequence causes the
analyzer to prompt for the
expected H
2
S span
concentration.
M-P CAL
H2S SPAN CONC: 450.0 Conc
0 0 0
4 5 0 .0
ENTR
EXIT
M-P CAL RANGE = 500.000 PPB
H2S
=XXX.X
< TST TST > ZERO
CONC
EXIT
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST >
CAL
SETUP
The H
2
S span concentration
values automatically default
to
450.0
ppb.
To change this value to the
actual concentration of the
span gas, enter the number
by pressing the key under
each digit until the expected
value appears.
The span gas concentration
should always be 90% of the
selected reporting range
EXAMPLE
Reporting range = 800 ppb
Span gas conc.= 720 ppb
EXIT
ignores the new setting
and returns to the previous
display.
ENTR
accepts the new setting
and returns to the
previous display..
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...