Calibration Procedures
Teledyne ML – T101 Operation Manual
180
Table 6-3. AutoCal Attribute Setup Parameters
Attribute Name
ACTION
Timer Enabled
Turns on the Sequence timer
Starting Date
Sequence will operate on Starting Date
Starting Time
Sequence will operate at Starting Time
Delta Days
Number of days to skip between each sequence
Delta Time
Incremental delay on each Delta Day that the sequence starts.
Duration
Duration of the sequence in minutes
Calibrate
Enable to do dynamic zero/span calibration, disable to do a cal check only.
This must be set to OFF for units used in US EPA applications and with IZS
option installed.
NOTE
The programmed STARTING_TIME must be a minimum of 5 minutes later than the real
time clock (See Section 4.4.6 for setting real time clock).
NOTE
Avoid setting two or more sequences at the same time of the day. Any new sequence
which is initiated whether from a timer, the COM ports, or the contact closure inputs will
override any sequence which is in progress.
NOTE
If at any time an illegal entry is selected (Example: Delta Days > 367) the ENTR button
will disappear from the display.
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...