Teledyne ML – T101 Operation Manual
EPA Protocol Calibration
195
7.5.
DYNAMIC MULTIPOINT SPAN CALIBRATION
Dynamic calibration involves introducing gas samples of known concentrations to an
instrument in order to record the instrument’s performance at a predetermined sensitivity
and to derive a calibration relationship. A minimum of three reference points and one zero
point uniformly spaced covering 0 to 90 percent of the operating range are recommended
to define this relationship.
The analyzer's recorded response is compared with the known concentration to derive the
calibration relationship.
To perform a precision check, the instrument set up, sources of zero air and sample gas
should conform to those described in Section 6.2.
Follow the procedures described in section 6.2 for calibrating the zero points.
For each mid point:
SPAN CAL M
RANGE = 500.0 PPB
SO
2
X.XXX
< TST TST >
ZERO
SPAN
CONC
EXIT
Wait until
SO2 STB
falls
below
0.5 ppb.
This may take
several minutes.
SAMPLE
SO2 STB =X.XXX PPB
SO2 =X.XXX
< TST TST > CAL
CALZ
CALS
SETUP
Record the SO
2
reading as
displayed on the
instrument’s front
panel
Press
EXIT
to
Return to the
Main
SAMPLE
Display
ACTION:
Allow Calibration Gas diluted to
proper concentration for
Midpoint N+1 to enter the sample
port
ACTION:
Allow calibration gas diluted to proper concentration for
Midpoint N to enter the sample port
SAMPLE* RANGE = 500.000 PPB SO2 =X.XXX
< TST
TST
>
CAL
SETUP
SAMPLE
SO2 STB =X.XXX PPB
SO2 =X.XXX
< TST TST >
CAL
SETUP
Set the Display to show the
SO2 STB
test function.
This function calculates the
stability of the SO
2
measurement
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...