Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485)
A-16
Setup Variable
Numeric
Units
Default
Value
Value Range
Description
RS232_MODE
BitFlag
0
0–65535
RS-232 COM1 mode flags.
Add values to combine flags.
1 = quiet mode
2 = computer mode
4 = enable security
16 = enable Hessen protocol
5
32 = enable multi-drop
64 = enable modem
128 = ignore RS-232 line
errors
256 = disable XON / XOFF
support
512 = disable hardware
FIFOs
1024 = enable RS-485 mode
2048 = even parity, 7 data
bits, 1 stop bit
4096 = enable command
prompt
BAUD_RATE —
115200
300,
1200,
2400,
4800,
9600,
19200,
38400,
57600,
115200
RS-232 COM1 baud rate.
Enclose value in double
quotes (“) when setting from
the RS-232 interface.
MODEM_INIT
—
“AT Y0 &D0
&H0 &I0
S0=2 &B0
&N6 &M0 E0
Q1 &W0”
Any
character in
the allowed
character
set. Up to
100
characters
long.
RS-232 COM1 modem
initialization string. Sent
verbatim plus carriage return
to modem on power up or
manually. Enclose value in
double quotes (“) when
setting from the RS-232
interface.
RS232_MODE2
BitFlag
0
0–65535
RS-232 COM2 mode flags.
(Same settings as
RS232_MODE.)
07266B DCN6485
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...