Teledyne ML – T101 Operation Manual
Operating Instructions
115
5. In the instrument’s SETUP>MORE>COMM>COM2 menu, make the following settings:
Baud Rate: 115200
COM2 Mode Settings:
Quiet Mode
ON
Computer Mode
ON
MODBUS RTU
OFF
MODBUS ASCII
OFF
E,8,1 MODE
OFF
E,7,1 MODE
OFF
RS-485 MODE
OFF
SECURITY MODE
OFF
MULTIDROP MODE
OFF
ENABLE MODEM
OFF
ERROR CHECKING
ON
XON/XOFF HANDSHAKE OFF
HARDWARE HANDSHAKE OFF
HARDWARE FIFO
ON
COMMAND PROMPT
OFF
6. Next, configure your communications software, such as APICOM. Use the COM port
determined in Step 4 and the baud rate set in Step 5. The figures below show how
these parameters would be configured in the Instrument Properties window in
APICOM when configuring a new instrument. See the APICOM manual (PN 07463)
for more details.
NOTE
USB configuration requires that the baud rates of the instrument and the PC match; check the PC baud
rate and change if needed.
Using the USB port disallows use of the rear panel COM2 port except for multidrop communication.
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...