Operating Instructions
Teledyne ML – T101 Operation Manual
76
In order to compensate for CPU clocks which run fast or slow, there is a variable to speed
up or slow down the clock by a fixed amount every day. To change this variable, press:
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL
SETUP
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
3x
EXIT
returns
to the main
SAMPLE
display
SETUP X.X 0 ) MEASURE_MODE=H2S
NEXT
JUMP
EDIT PRNT EXIT
Enter sign and number of seconds per
day the clock gains (-) or loses (+).
SETUPX.X 1 ) DAS_HOLD_OFF=15.0 Minutes
PREV
NEXT
JUMP
EDIT PRNT EXIT
SETUP X.X
8) CLOCK_ADJ=0 Sec/Day
PREV JUMP
EDIT
PRNT EXIT
SETUP X.X
CLOCK_ADJ:
0 Sec/Day
+ 0 0
ENTR
EXIT
SETUP X.X
8) CLOCK_ADJ=0 Sec/Day
PREV NEXT JUMP
EDIT PRNT
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
DIAG
EXIT
Continue to press
NEXT
until …
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...