Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485)
A-28
Signal Name
Bit or Channel
Number
Description
DARK_SHUTTER
10
0 = close dark shutter
1 = open
FLUSH_VALVE
4
11
0 = open flow restrictor bypass (flush) valve
1 = close
12–15
Spare
AUX board analog inputs, default I
2
C address 30 hex
PMT_SIGNAL
7
0 (register number)
PMT detector
UVLAMP_SIGNAL
7
1
UV
lamp
intensity
NORM_PMT_SIGNAL
7
2
Normalized PMT detector
PMT_TEMP
7
3 PMT
temperature
HVPS_VOLTAGE
7
4
HV power supply output
PMT_DARK
7
5
PMT reading during dark cycles
LAMP_DARK
7
6
Lamp reading during dark cycles
AGND_DARK
7
7
AGND reading during dark cycles
AGND_LIGHT
7
8
AGND reading during light cycles
VREF_DARK
7
9
VREF4096 reading during dark cycles
VREF_LIGHT
7
10
VREF4096 reading during light cycles
Rear board primary MUX analog inputs
PMT_SIGNAL 0
PMT
detector
HVPS_VOLTAGE
1
HV power supply output
PMT_TEMP 2 PMT
temperature
UVLAMP_SIGNAL 3
UV
lamp
intensity
4
Temperature
MUX
PHOTO_ABS
5
Pre-amplified UV lamp intensity
OXY_FLOW
2
6
Oxygenator flow rate
SAMPLE_PRESSURE 7
Sample
pressure
TEST_INPUT_8
8
Diagnostic test input
REF_4096_MV
9
4.096V reference from MAX6241
SAMPLE_FLOW 10
Sample
flow
rate
TEST_INPUT_11
11
Diagnostic test input
CONV_TEMP
1
12
Converter
temperature
13
Spare (thermocouple input?)
14
DAC
MUX
REF_GND 15 Ground
reference
Rear board temperature MUX analog inputs
BOX_TEMP 0 Internal
box
temperature
RCELL_TEMP 1
Reaction
cell
temperature
IZS_TEMP 2 IZS
temperature
3
Spare
07266B DCN6485
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...