Operating Instructions
Teledyne ML – T101 Operation Manual
100
4.6.7.
PRESSURE CALIBRATION
A sensor at the exit of the sample chamber continuously measures the pressure of the
sample gas. This data is used to compensate the final H
2
S concentration calculation for
changes in atmospheric pressure when the instrument’s TPC feature (see Section 10.7.3)
is turned on and is stored in the CPU’s memory as the test function
PRES
(also viewable
via the front panel).
To cause the analyzer to measure and record a value for
PRES
, press:
Exit at
any time
to return
to main
the
SETUP
menu
DIAG
SIGNAL I / O
NEXT
ENTR
EXIT
Repeat Pressing
NEXT
until . .
.
DIAG
PRESSURE CALIBRATION
PREV NEXT
ENTR
EXIT
DIAG PCAL
ACTUAL PRES :27.20 IN-HG-A
2 7 .2 0 ENTR EXIT
E
NTR
accepts the
new value
EXIT
ignores the new
value
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL
SETUP
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
Adjust these values until the
displayed pressure equals the
pressure measured by the
independent pressure meter.
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...