Teledyne ML – T101 Operation Manual
Operating Instructions
69
4.4.4.4.
Single Range Mode (SNGL)
The default range mode for the analyzer is single range, in which all analog concentration
outputs are set to the same reporting range. This reporting range can be set to any value
between 5.0 ppb and 20 000 ppb.
While the two outputs always have the same reporting range, the span and scaling of their
electronic signals may also be configured for different differently (e.g.,
A1
= 0-10 V;
A2
= 0-0.1 V).
To select
SNGLE
range mode and to set the upper limit of the range, press:
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP C.3
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP C.3
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
SETUP C.3
RANGE MODE:
SNGL
SNGL
IND AUTO
ENTR EXIT
EXIT
x 2 returns
to the main
SAMPLE
display
SETUP C.3
RANGE MODE:
SNGL
SNGL
IND AUTO
ENTR
EXIT
SETUP C.3
RANGE CONTROL MENU
MODE
SET
UNIT
EXIT
SETUP C.3
RANGE:
500.0 Conc
0 0
5
0 0 .0
ENTR
EXIT
SETUP C.3
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL
SETUP
NOTE
On analyzers with the multigas option activated (see Sections 4.5.1 and 6.8) the concentration value will
switch back and forth between from “H2S=XXX.X” to “SO2=XXX.X” depending on which gas is currently
being measured.
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...