Operating Instructions
Teledyne ML – T101 Operation Manual
70
4.4.4.5.
Independent Range Mode (IND)
Selecting independent range mode allows the
A1
and
A2
outputs to be configured with
different measurement ranges. The analyzer software calls these two ranges
LOW
and
HIGH
. The
LOW
range setting corresponds with the analog output labeled
A1
on the
rear panel of the instrument. The
HIGH
range setting corresponds with the
A2
output.
While the software names these two ranges
LOW
and
HIGH
, they do not have to be
configured that way.
Also, in this mode the
RANGE
Test function displayed on the front panel during
SAMPLE
mode will be replaced by two separate functions,
RANGE1
&
RANGE2
.
•
LOW range = RANGE1 = Range value for output A1 = 0-1500 ppb
H
2
S.
•
HIGH range = RANGE2 = Range value for output A2 = 0-500 ppb
H
2
S.
For T101’s configured to measure both SO
2
and H
2
S in multigas measurement mode:
•
LOW range = RANGE1 = Range value for output A1= 0-1500 ppm
SO
2
.
•
HIGH range = RANGE2 = Range value for output A2 =0-1000 ppm
H
2
S.
To select the independent reporting range mode and set the upper measurement limits for
the two outputs, press:
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
SETUP X.X
RANGE MODE:
SNGL
SNGL
IND
AUTO
ENTR EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET
UNIT
EXIT
SETUP X.X
HIGH RANGE:
500.0 Conc
0 0 5
0 0 .0
ENTR
EXIT
SETUP X.X
LOW RANGE
:
500.0 Conc
0 0
1
0 0 .0
ENTR
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
EXIT
Returns
to the Main
SAMPLE
Display
SETUP X.X
RANGE MODE:
DUAL
SNGL
IND AUTO
ENTR
EXIT
Toggle the
buttons under
each digit to set
the upper limit
of each range.
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL
SETUP
NOTE
In INDEPENDENT range mode the two reporting ranges have separate slopes and offsets
for computing concentration and MUST be independently calibrated.
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...