Getting Started
Teledyne ML – T101 Operation Manual
42
PMT
EXHAUST THROUGH OUTER
LAYER OF KICKER
RE
ACT
IO
N
CE
L
L
P
URG
E
FLOW
SENSOR
FLOW / PRESSURE
SENSOR PCA
SAMPLE
FILTER
INSTRUMENT CHASSIS
EXHAUST GAS
OUTLET
V
ACUUM
M
ANI
F
O
L
D
SAMPLE GAS
INLET
KICKER EXHAUST TO PUMP
PUMP
FLOW
CONTROL
ASSY
SAMPLE
PRESSURE
SENSOR
UV
LAMP
HYDROCARBON
SCRUBBER
(KICKER)
SPAN GAS INLET
ZERO AIR INLET
SAMPLE
CHAMBER
COM
NC
NO
SO
2
Scrubber
MOLYBDENUM
CONVERTER
SO
2
H
2
S
H
2
S / SO
2
MODE VALVE
Gas Flow when multigas version of
Analyzer is measuring SO
2
.
Figure 3-13. Pneumatic Diagram of the T101 Standard Configuration
Table 3-8. H
2
S – SO
2
Switching Valve Operating States
GAS
MODE
CONDITION OF H
2
S –SO
2
SWITCHING
VALVE
VALVE PORT
CONNECTION
H
2
S
Open to SO
2
Scrubber and Molybdenum
Converter
COM
NO
SO
2
Open to directly to Sample Chamber. Bypasses
SO
2
Scrubber and Molybdenum Converter
COM
NC
H
2
S –SO
2
Switches between above two states every 10
minutes.
- -
1.
Attach the 1/4" exhaust line to the EXHAUST port of the analyzer and to
the inlet port of the pump.
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...