Teledyne ML – T101 Operation Manual
Getting Started
51
Table 3-12. Possible Warning Messages at Start-Up
WARNING MESSAGE
MEANING
ANALOG CAL WARNING
The instrument’s A/D circuitry or one of its analog outputs is not
calibrated.
BOX TEMP WARNING
The temperature inside the T101 chassis is outside the specified limits.
CANNOT DYN SPAN
Remote span calibration failed while the dynamic span feature was set to
active
CANNOT DYN ZERO
Remote zero calibration failed while the dynamic zero feature was set to
active
CONFIG INITIALIZED
Configuration was reset to factory defaults or was erased.
SHUTTER WARNING
Dark offset above limit specified indicating that too much stray light is
present in the sample chamber.
DATA INITIALIZED
DAS data storage was erased.
HVPS WARNING
High voltage power supply for the PMT is outside of specified limits.
IZS TEMP WARNING
On units with IZS options installed: The permeation tube temperature is
outside of specified limits.
PMT DET WARNING
PMT detector output outside of operational limits.
PMT TEMP WARNING
PMT temperature is outside of specified limits.
RCELL TEMP WARNING
Sample chamber temperature is outside of specified limits.
REAR BOARD NOT DET
The CPU is unable to communicate with the motherboard.
RELAY BOARD WARN
The firmware is unable to communicate with the relay board.
SAMPLE FLOW WARN
The flow rate of the sample gas is outside the specified limits.
SAMPLE PRESS WARN
Sample pressure outside of operational parameters.
SYSTEM RESET
The computer was rebooted.
UV LAMP WARNING
The UV lamp intensity measured by the reference detector reading too
low or too high
To view and clear warning messages:
SAMPLE
HVPS WARNING SO2 = 0.00
TEST
CAL
MSG
CLR
SETUP
Press
CLR
to clear the displayed
message.
If more than one warning is active, the
next message is displayed.
SAMPLE
RANGE=500.000 PPB
SO2 = 0.00
< TST TST > CAL
MSG
CLR
SETUP
SAMPLE
SYSTEM RESET
SO2 = 0.00
TEST CAL
MSG
CLR
SETUP
TEST
suppresses the
warning messages
MSG
returns active warning
messages to the Param field.
NOTE:
If the warning message persists
after several attempts to clear it,
the message may indicate a
real problem and not an artifact
of the warm-up period
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...