Teledyne ML – T101 Operation Manual
Calibration Procedures
185
6.10.
CALIBRATION QUALITY
After completing one of the calibration procedures described above, it is important to
evaluate the analyzer’s calibration slope and offset
parameters. These values describe the
linear response curve of the analyzer. The values for these terms, both individually and
relative to each other, indicate the quality of the calibration. To perform this quality
evaluation, you will need to record the values of both test functions (Section 4.2.1 or
Appendix A-3), all of which are automatically stored in the DAS channel
CALDAT
for
data analysis, documentation and archival.
Make sure that these parameters are within the limits listed in Table 6-5.
Table 6-5. Example Auto-Cal Sequence
FUNCTION
MINIMUM VALUE
OPTIMUM VALUE
MAXIMUM VALUE
H2S SLOPE
1
-0.700
1.000
1.300
H2S OFFS
1
-50.0 mV
<100
250.0 mV
These values should not be significantly different from the values recorded on the Teledyne ML Final Test
and Validation Data sheet that was shipped with your instrument. If they are, refer to the troubleshooting
Section 9.
1
Shown as they appear when analyzer is in H
2
S mode. In SO
2
mode they appear as SO2 OFFS & SO2
SLOPE. In multigas mode, both versions are listed and should be checked
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...