Table 9-9. Example of UV Lamp Power Supply Outputs
UVLAMP_SIGNAL
ACTION TO BE TAKEN
3500mV±200mV.
No Action Required
> 4900mV at any time.
Adjust the UV reference detector potentiometer (refer to Figure 9-6)
until UVLAMP_SIGNAL reads approximately 3600mV before
continuing to adjust the lamp position.
>3700mV or < 3300mV Adjust the UV reference detector potentiometer (refer to Figure 9-6)
until UVLAMP_SIGNAL reads as close to 3500mV as possible.
.< 600mV
Replace the lamp.
Figure 9-5. Location of UV Reference Detector Potentiometer
5.
Finger-tighten the thumbscrew.
NOTE
DO NOT over-tighten the thumbscrew.
9.6.3.
REPLACING THE UV LAMP
1.
Turn off the analyzer.
27.
Disconnect the UV lamp from its power supply.
•
You can find the power supply connector by following the two, white
UV Lamp power supply wires from the lamp to the power supply.
28.
Loosen, but do not remove the two UV lamp bracket screws and the
large brass thumbscrew located on the shutter housing (refer to Figure
9-4) so that the lamp can be moved.
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...