Operating Instructions
Teledyne ML – T101 Operation Manual
66
4.4.2.
SETUP – ACAL: AUTO CALIBRATION
Used to set up and operate the internal valve options if installed. Section 6 provides
details.
4.4.3.
SETUP – DAS: DATA ACQUISITION
Used to set up the data acquisition system and record data.
4.4.4.
SETUP – RANGE: ANALOG OUTPUT REPORTING RANGE
CONFIGURATION
4.4.4.1.
Available Analog Output Signals
The analyzer has three active analog output signals, accessible through a connector on the
rear panel.
ANALOG OUT
A1 A2 A3 A4
+ - + - + - + -
H
2
S/SO
2
concentration outputs
HIGH range when
DUAL mode is selected
Not Used
Test Channel
LOW range when
DUAL mode is selected
Figure 4-3. Analog Output Connectors Defined
All three outputs can be configured either at the factory or by the user for full scale
outputs of 0.1 VDC, 1VDC, 5VDC or 10VDC. Additionally
A1
and
A2
may be equipped
with optional 0-20 mADC current loop drivers and configured for any current output
within that range (e.g. 0-20, 2-20, 4-20, etc.). The user may also adjust the signal level
and scaling of the actual output voltage or current to match the input requirements of the
recorder or data logger (See Sections 4.6.3.3 and 4.6.3.5).
In its basic configuration, the
A1
and
A2
channels of the T101 output a signal that is
proportional to the H
2
S concentration of the sample gas. Several operating modes are
available which allow them to be slaved together (
SNGL
Mode, see Section 4.4.4.4 or
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...