Calibration Procedures
Teledyne ML – T101 Operation Manual
172
Step Three:
Perform the calibration or calibration check according to the following flow
chart:
Wait until
H2S
STB
falls below
0.5 ppb. This may
take several
minutes.
ZERO CAL M
H2S STB=XXX.X PPB
H2S
=XXX.X
< TST TST >
ZERO
CONC EXIT
EXIT
returns to the
main
SAMPLE
display
ZERO CAL M H2S STB=XXX.X PPB
H2S
=XXX.X
< TST TST > ENTR CONC
EXIT
Pressing
ENTR
changes the calibration of the instrument.
If either the
ZERO
or
SPAN
button fails to
appear, see Chapter 11
for troubleshooting tips.
Analyzer enters
ZERO
CAL
mode
.
EXIT
returns the unit to
SAMPLE
mode without
changing the calibration
values.
ZERO CAL M H2S STB=XXX.X PPB
H2S =X.XXX
< TST TST > ZERO CONC
EXIT
ZERO CAL M H2S STB=XXX.X PPB H2S
=XXX.X
< TST TST > CAL
CALZ
CALS
SETUP
SPAN CAL M
H2S STB=XXX.X PPB
H2S
=XXX.X
< TST TST >
SPAN
CONC
EXIT
SPAN CAL M H2S STB=XXX.X PPB
H2S
=XXX.X
< TST TST >
ENTR
CONC
EXIT
EXIT
returns to the
SAMPLE
mode without
changing the calibration
values.
SPAN CAL M H2S STB=XXX.X PPB
H2S
=XXX.X
< TST TST >
SPAN
CONC
EXIT
ACTION:
Allow zero gas to enter the sample port at the
rear of the instrument.
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST
TST >
CAL
CALZ
CALS
SETUP
SAMPLE
H2S STB=XXX.X PPB
H2S
=XXX.X
< TST TST > CAL
CALZ
CALS
SETUP
Scroll the display to the
H2S
STB
test function. This function
calculates the stability of the
H
2
S measurements.
The value of
H2S
STB
may jump
significantly. Wait
until it falls below
0.5
ppb. This may take
several minutes.
Pressing
ENTR
changes the calibration of the instrument.
Analyzer enters
SPAN
CAL
Mode.
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...