Teledyne ML – T101 Operation Manual
Operating Instructions
63
The default status of the calibration password is
OFF
. To enable the calibration password
press:
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SAMPLE RANGE = 500.0 PPB H2S =X.XXX
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS
CLK MORE
EXIT
SETUP X.X
CAL. PASSWORD ENABLE: OFF
OFF
ENTR EXIT
SETUP X.X
PASSWORD ENABLE:
ON
ON
ENTR EXIT
SETUP X.X
PASSWORD ENABLE: ON
ON
ENTR
EXIT
ENTR
accepts
the change
EXIT
ignores
the change
Toggles
password
status On/Off
ENTR
accepts
displayed
password value
EXIT
returns to
SAMPLE
display
CAL. PASSWORD
default state is
OFF
If the calibration password (
101
) is enabled, the following menu button sequence will be
required to enter one of the calibration modes:
SAMPLE
ENTER SETUP PASS : 0
0 0 0
ENTR
EXIT
SAMPLE RANGE = 500.0 PPB H2S =X.XXX
< TST TST >
CAL CALZ CALS
SETUP
M-P CAL RANGE = 500.0 PPB H2S =X.XXX
< TST TST > ZERO
CONC
EXIT
Prompts
password
number
SAMPLE
ENTER SETUP PASS : 0
1 0 1
ENTR
EXIT
Press
individual
buttons to set
101
Continue calibration process …
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...