Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485)
A-23
Table A-3: Test Measurements
TEST Measurement
Message Text
DESCRIPTION
RANGE RANGE=500.0
PPB
3
D/A range in single or auto-range
modes.
RANGE1 RANGE1=500.0
PPB
3
D/A #1 range in independent range
mode.
RANGE2 RANGE2=500.0
PPB
3
D/A #2 range in independent range
mode.
RANGE3
5
RANGE3=500.0
PPB
3
D/A #3 range in independent range
mode.
STABILITY
SO2 STB=11.4 PPB
3
Concentration
stability
#1.
STABILITY2
8
SO2 STB2=6.3 PPB
3
Concentration
stability
#2.
RESPONSE
2
RSP=1.11(0.00) SEC
Instrument response. Length of
each signal processing loop. Time in
parenthesis is standard deviation.
OXYFLOW
6
OXY FLOW=150 CC/M
Oxygenator flow rate
SAMPFLOW
SAMP FL=700 CC/M
Sample flow rate.
SAMPPRESS
PRES=29.9 IN-HG-A
Sample pressure.
PMTDET
PMT=762.5 MV
Raw PMT reading.
NORMPMTDET
NORM PMT=742.9 MV
PMT reading normalized for
temperature, pressure, auto-zero
offset, but not range.
UVDET
UV LAMP=3457.6 MV
UV lamp reading.
STABILITYUV
9
UV STB=5.607 MV
UV lamp stability reading.
LAMPRATIO
LAMP RATIO=100.0 %
UV lamp ratio of current reading
divided by calibrated reading.
STRAYLIGHT
STR. LGT=0.1 PPB
Stray light offset.
DARKPMT
DRK PMT=19.6 MV
PMT dark offset.
DARKLAMP
DRK LMP=42.4 MV
UV lamp dark offset.
SO2SLOPE
SO2 SLOPE=1.000
Slope for current range, computed
during zero/span calibration.
SO2OFFSET
SO2 OFFS=0.0 MV
Offset for current range, computed
during zero/span calibration.
H2SSLOPE
H2S SLOPE=1.000
Slope for current range, computed
during zero/span calibration.
H2SOFFSET
H2S OFFS=0.0 MV
Offset for current range, computed
during zero/span calibration.
TRSSLOPE
5
TRS SLOPE=1.000
Slope for current range, computed
during zero/span calibration.
TRSOFFSET
5
TRS OFFS=0.0 MV
Offset for current range, computed
during zero/span calibration.
HVPS
HVPS=650 VOLTS
High voltage power supply output.
RCELLDUTY
2
RCELL ON=0.00 SEC
Reaction cell temperature control
duty cycle.
RCELLTEMP
RCELL TEMP=52.1 C
Reaction cell temperature.
BOXTEMP
BOX TEMP=35.5 C
Internal chassis temperature.
07266B DCN6485
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...