Teledyne ML – T101 Operation Manual
Operating Instructions
129
LIST OF PARAMETERS
NAME:
PNUMTC
EVENT:
ATIMER
REPORT PERIOD:
001:00:00
NO. OF RECORDS:
360
RS-232 REPORT:
OFF
CHANNEL ENABLED:
ON
CAL HOLD OFF:
OFF
NAME:
CALDAT
EVENT:
SLPCHG
NO. OF RECORDS:
200
RS-232 REPORT:
OFF
CHANNEL ENABLED:
ON
CAL HOLD OFF:
OFF
PARAMETER:
SMPLFLW
MODE:
AVG
PRECISION:
1
STORE NUM SAMPLES
OFF
PARAMETER:
SMPLPRS
MODE:
AVG
PRECISION:
1
STORE NUM SAMPLES
OFF
PARAMETER:
S2OFS1
MODE:
INST
PRECISION:
1
STORE NUM SAMPLES
OFF
PARAMETER:
S2SLP1
MODE:
INST
PRECISION:
3
STORE NUM SAMPLES
OFF
PARAMETER:
S2ZSC1
MODE:
INST
PRECISION:
1
STORE NUM SAMPLES
OFF
PARAMETER:
S2CONC1
MODE:
AVG
PRECISION:
1
STORE NUM SAMPLES
OFF
PARAMETER:
S2CONC1
MODE:
AVG
PRECISION:
1
STORE NUM SAMPLES
OFF
NAME:
CONC
EVENT:
ATIMER
REPORT PERIOD:
000:00:05
NO. OF RECORDS:
4032
RS-232 REPORT:
OFF
CHANNEL ENABLED:
ON
CAL HOLD OFF:
ON
LIST OF CHANNELS
Figure 4-11. Default DAS Channels Setup
These default Data Channels can be used as they are, or they can be customized from the
front panel to fit a specific application. They can also be deleted to make room for custom
user-programmed Data Channels.
Appendix A-5 lists the firmware-specific DAS configuration in plain-text format. This
text file can either be loaded into APICOM and then modified and uploaded to the
instrument or can be copied and pasted into a terminal program to be sent to the analyzer.
NOTE
Sending an DAS configuration to the analyzer through its COM ports will replace the
existing configuration and will delete all stored data. Back up any existing data and the
DAS configuration before uploading new settings.
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...