Operating Instructions
Teledyne ML – T101 Operation Manual
124
4.7.12.
COM PORT TESTING
The serial ports can be tested for correct connection and output in the COMM menu. This
test sends a string of 256 ‘w’ characters to the selected COM port. While the test is
running, the red LED on the rear panel of the analyzer should flicker.
To initiate the test press the following button sequence.
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL
SETUP
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
Select which
COM port to
test.
SETUP X.X
COMMUNICATIONS MENU
ID INET
COM1
COM2
EXIT
SETUP X.X
COM1 : TEST PORT
<SET
TEST
EXIT
SETUP X.X
TRANSMITTING TO COM1
<SET
TEST
EXIT
SETUP X.X
COM1 MODE:0
SET>
EDIT
EXIT
SETUP X.X COM1 BAUD RATE:19200
<SET
SET>
EDIT
EXIT
EXIT returns to
COMM menu
Test runs
automatically
4.8.
USING THE DATA ACQUISITION SYSTEM (DAS )
The T101 analyzer contains a flexible and powerful, internal data acquisition system
(DAS ) that enables the analyzer to store concentration and calibration data as well as a
host of diagnostic parameters. The DAS of the T101 can store several months of data,
depending on how it is configured. The data are stored in non-volatile memory and are
retained even when the instrument is powered off. Data are stored in plain text format for
easy retrieval and use in common data analysis programs (such as spreadsheet-type
programs).
The DAS is designed to be flexible. Users have full control over the type, length and
reporting time of the data. The DAS permits users to access stored data through the
instrument’s front panel or its communication ports. Using APICOM, data can even be
retrieved automatically to a remote computer for further processing.
The principal use of the DAS is logging data for trend analysis and predictive diagnostics,
which can assist in identifying possible problems before they affect the functionality of
the analyzer. The secondary use is for data analysis, documentation and archival in
electronic format.
To support the DAS functionality, Teledyne ML offers APICOM, a program that
provides a visual interface for remote or local setup, configuration and data retrieval of
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...