Operating Instructions
Teledyne ML – T101 Operation Manual
80
4.5.1.
SETTING THE GAS MEASUREMENT MODE
If the switching valves software is activated, the T101 can be set to one of three gas
measurement modes:
•
H
2
S
The sample gas stream is stripped of any ambient SO
2
by a special
chemical scrubber, then passed through a catalytic converter that changes
the H
2
S present into SO
2
which is then measured using the UV
Fluorescence method
•
SO
2
The sample gas stream bypasses the SO
2
Scrubber and catalytic
converter allowing the only ambient SO
2
to be measured.
•
H
2
S – SO
2
The switching valve alternates the gas stream between the two paths at
regular intervals allowing the instrument to measure both gases.
To select one of these three measurement modes (see Section 10.3.2 for additional
details), press:
SETUP X.X
0 ) MEASURE_MODE=H2S
NEXT
JUMP
EDIT
PRNT EXIT
EXIT
ignores the new
setting.
ENTR
accepts the
new setting.
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
DIAG
EXIT
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL
SETUP
SETUP X.X
MEASURE MODE: H2S
PREV
ENTR
EXIT
SETUP X.X
MEASURE MODE: H2S-SO2
PREV
NEXT
ENTR
EXIT
SETUP X.X
MEASURE MODE: SD2
NEXT
ENTR
EXIT
H
2
S mode is the
default mode.
Press the
PREV
and
NEXT
buttons
to scroll among gas
measurement
mode choices.
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...