Principles Of Operation
257
10.3.2.
MULTIGAS MEASUREMENT & H
2
S
SO
2
SWITCHING
VALVE
When activated for operation the multigas measurement mode allows the instrument to
measure either or both H
2
S or SO
2
via a Teflon
®
switching valve. This valve, under
CPU control via the I
2
C buss and the relay board, directs the sample gas stream either
through the SO
2
scrubber and H
2
S
SO
2
converter (H
2
S measurement mode) or
directly to the sample chamber bypassing the H
2
S
SO
2
converter, allowing the
analyzer to measure SO
2
. The cycle for this operation is
Table 10-1. T101 Multigas Valve Cycle-Phases
Gas Mode
H
2
S
SO
2
Valve
Status
Default
Time
Settings
Activity
H
2
S
Gas stream directed
through scrubber and
converter
0 – 3
minutes
Wait period. Ensures sample chamber has
been flushed of previous gas.
3 – 10 m
Analyzer measures florescence in sample
chamber
SO
2
Gas stream bypasses
through scrubber and
converter
0 – 3
minutes
Wait period (dwell time). Ensures sample
chamber has been flushed of previous gas.
3 – 10 m
Analyzer measures florescence in sample
chamber
Cycle repeats every ~20Minuites
The timing of the above cycle is set by two variables (see Appendix A-2),
MEASURE_PERIOD,
which sets the total dwell time for each gas mode, and
MEASURE_DELAY
which sets the wait period before the instrument begins making
measurements after the gas mode has been switch.
10.3.3.
FLOW RATE CONTROL
The Model T101 uses a special flow control assembly located in the exhaust vacuum
manifold (Figure 10-7) to maintain a constant flow rate of the sample gas through the
instrument. This assembly consists of:
•
a critical flow orifice
•
two o-rings: Located just before and after the critical flow orifice, the o-
rings seal the gap between the walls of assembly housing and the critical
flow orifice.
•
a spring: Applies mechanical force needed to form the seal between the
o-rings, the critical flow orifice and the assembly housing.
10.3.3.1.
Critical Flow Orifice
The most important component of this flow control assembly is the critical flow orifice.
Critical flow orifices are a remarkably simple way to regulate stable gas flow rates. They
operate without moving parts by taking advantage of the laws of fluid dynamics. By
restricting the flow of gas though the orifice, a pressure differential is created. This
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...