Principles Of Operation
Teledyne ML - T101 Operation Manual
256
10.3.1.
SAMPLE GAS FLOW
The flow of gas through the T101 UV Fluorescence H
2
S Analyzer is created by a small
internal pump that pulls air though the instrument.
PMT
EXHAUST TO OUTER
LAYER OF
KICKER
RE
ACT
IO
N C
E
L
L
P
URG
E
FLOW
SENSOR
FLOW / PRESSURE
SENSOR PCA
SAMPLE
FILTER
INSTRUMENT CHASSIS
EXHAUST GAS
OUTLET
V
AC
UU
M
MANI
FO
LD
SAMPLE GAS
INLET
KICKER EXHAUST TO PUMP
PUMP
FLOW
CONTROL
ASSY
SAMPLE
PRESSURE
SENSOR
UV
LAMP
HYDROCARBON
SCRUBBER
(KICKER)
SPAN GAS INLET
ZERO AIR INLET
SAMPLE
CHAMBER
2
1
3
SO
2
Scrubber
MOLYBDENUM
CONVERTER
SO
2
H
2
S
H
2
S / SO
2
MODE VALVE
Gas Flow when multigas version of
Analyzer is measuring SO
2
.
Figure 10-7. T101 Gas Flow and Location of Critical Flow Orifice
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...