Calibration Procedures
179
The analyzer will ask to have the
GAS TYPE
specified at the beginning of the process as
follows:
Continue Calibration as per
Standard Procedure
SAMPLE H2S STB =XXX.X PPB
H2S
=XXX.X
< TST TST >
CAL
CALZ CALS
SETUP
ZERO CAL M
H2S STB =XXX.X PPB
H2S
=XXX.X
< TST TST >
ZERO
SPAN
CONC
EXIT
SAMPLE
GAS TO CAL: H2S
SO2 H2S
ENTR
SETUP
SAMPLE
GAS TO CAL: SO2
LOW HIGH
ENTR
SETUP
Wait until
H2S
STB
falls below
0.5 ppb. This may
take several
minutes.
Once this selection is made, the calibration procedure continues as previously described.
The other gas may be calibrated by starting over from the main
SAMPLE
display.
6.9.
AUTOMATIC CALIBRATION/CHECKS (AUTOCAL)
The AutoCal system allows unattended, periodic operation of the zero/span valve options
by using the analyzer’s internal time of day clock. AutoCal operates by executing user-
defined sequences to initiate the various calibration modes of the analyzer and to open
and close valves appropriately. It is possible to program and run up to three separate
sequences (
SEQ1
,
SEQ2
and
SEQ3
). Each sequence can operate in one of three modes
or be disabled.
Table 6-2. AutoCal Modes
MODE
ACTION
DISABLED
Disables the sequence
ZERO
Causes the sequence to perform a zero calibration or check
ZERO-SPAN
Causes the sequence to perform a zero and span concentration calibration or
check
SPAN
Causes the sequence to perform a span concentration calibration or check
Each mode has seven parameters that control operational details of the sequence
(Table 6-3).
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...