
Processing
5-16
20077/021 US
X
TRA
Operator’s Manual
Emptying the RBC Line Without Ending the Case
From the End of Case Screen, it is possible to empty any blood remaining in the bowl or
empty line. To do so, touch the Empty button.
The system enters the Empty phase, and the Empty Phase of the Processing Screen is displayed on
the touch screen (refer to “The Empty Phase” on page 5-10).
Closing the End of Case Screen Without Ending the Case
If you decide you do not want to end the current case, the End of Case Screen may be closed and
further cycles run in the current case.
To close the End of Case Screen without ending the current case, touch the Close button.
The Ready Screen will be displayed on the touch screen (refer to “Processing a Cycle in Automatic or 1
Touch Mode” on page 5-7).
Retaining Case Data Between Power Cycles
If the X
TRA
system is powered OFF (through the ON/OFF switch or a power failure) before it has been
unloaded, then the case data and any modifications to the protocols are retained in memory and may
be resumed at the next power on.
To resume such an interrupted case, perform the following actions:
1.
Switch on the system using the ON/OFF switch at the rear of the machine to enter the Setup
Screen.
2.
From the Setup Screen touch the “Retain” button. (Do not touch the Load button.)
This initiates the usual auto-load procedure, except case data is not reset and the pump loop tubing is
assumed to be already loaded. Once the auto-load procedure has completed, the Ready Screen will be
displayed as usual.
Note:
The “Retain” button is not present on the Setup Screen under any of these conditions:
• The operator explicitly ended the case by touching the Unload button from the End of Case
Screen,
• The operator modified any of the machine settings through the Settings tab of the Menu Screen
(made exception for the Vacuum pump Intra/Post-op mode switch),
or
• A memory failure or initialization occurred.
WARNING
Repeating this operation may cause air to enter the RBC bag with
serious risks to the health of the patient. The air in the bag should
be removed before reinfusion.
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...