System Description
3-12
20077/021 US
X
TRA
Operator’s Manual
Processing Pump
The processing pump (see Figure 3-16) moves fluids into and out of the centrifuge bowl while
providing high flow rates with minimal hemolysis. In the autotransfusion protocols, the processing
pump can be operated at rates of 25 to 1,000 ml/min. in 25 ml/min. increments. In both the PPP and
PRP protocols, the pump can be operated at rates of 10 to 100 ml/min. in 10 ml/min. increments.
Figure 3-16 Visible Parts of the X
TRA
Processing Pump
The processing pump can be unfixed from its seat for cleaning purposes, as described in
Air Detector
The air detector is located on the top console between the processing pump and the centrifuge bowl
(see Figure 3-17).
Figure 3-17 Location of the Air Detector
The air detector is an ultrasonic device that detects air in the tubing from the pump to the centrifuge
bowl. It is responsible for detecting the end point of the Empty phase and detecting when the
collection reservoir or wash bag becomes empty.
2
1 Pump rotor
2 Pump handle
1
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...