Processing
5-8
20077/021 US
X
TRA
Operator’s Manual
Once the X
TRA
is loaded, the Ready Screen is displayed (see Figure 5-14). The system is now ready to
begin a processing cycle.
Figure 5-14 Ready Screen
For a detailed explanation of the
Ready Screen, read “Ready
Screen” on page 5-38.
1.
Touch the Start button to begin the Fill phase.
If the Autostart automation is ON (it is always ON in the 1 Touch mode and available as an option
in the Automatic mode), then it is not necessary to touch the Start button. The Fill phase will
begin processing automatically once the reservoir contains the programmed Autostart volume.
The Fill Phase
When the Fill phase begins (by touching the Start button or by the Autostart automation), the
Processing Screen is displayed on the touch screen, the reservoir clamp opens, and the centrifuge
begins accelerating. Once the centrifuge reaches its target speed, the pump will begin rotating and
filling the bowl with blood from the reservoir.
The Fill phase, like all processing phases, is monitored and controlled from the processing displet of
the Processing Screen (see Figure 5-15).
WARNING
Do not touch any moving parts of the centrifuge or pump. Injury may
result.
WARNING
A trained operator should be present at all times to monitor the X
TRA
system during processing. During operation, the X
TRA
should never be
left unattended. Unattended processing can lead to the development of
problems with the operation of the system and/or with the quality of the
end product. If the pump does not stop after the Stop button is pressed,
the operator must shut OFF the power.
CAUTION
Always close the centrifuge lid before starting any function of the machine
that uses the pump and/or centrifuge action to avoid the risk of touching any
1
1 Start button
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...