Installing the Disposables
4-8
20077/021 US
X
TRA
Operator’s Manual
Figure 4-6 Inserting the Centrifuge Bowl
10. Close the centrifuge arm by swinging it forward (see “B” in Figure 4-6).
11. Manually rotate the bowl to ensure that it rotates without eccentricity, by viewing from directly
above “shoulder” of bowl (see Figure 4-7). If the bowl does not rotate properly on its
perpendicular axis, remove the bowl and repeat installation.
Figure 4-7 Inserting the Bowl in Order to Avoid Wobbling
12. Line up the tubing so that the pump loop cartridge snaps into the notch provided and the
organizer is properly seated in its spot (see the left image of Figure 4-8). Ensure that the
color-coded tubings are positioned over the clamps and are not tangled.
WARNING
Do not attempt operation of the machine with improperly seated bowl.
Immediately stop operation of the system in case of unusually noisy bowl
rotation.
B
C
A
Look here
Bowl shoulder
Internal white
Bowl collar
Do not
look here
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...