Introduction and Safety
X
TRA
Operator’s Manual
20077/021 US
1-9
Storage and Transporting Warnings
1.
When not in use, the unit must be stored under the specified environment conditions in a
dust-free place, covered with the protective cover supplied.
2.
When the unit is not in use, switch off the main switch, disconnect the plug from the socket and
wind the power cord around the holders on the rear of the unit.
3.
The unit must be stored in a cool, dry and well ventilated place.
4.
When closing the lateral pins, ensure your fingers are kept away from the levers; otherwise you
expose yourself to a serious risk of injury.
Electrical Warnings
1.
Before connecting the unit, check that the mains socket is supplied at the voltage and frequency
shown on the label on the unit.
2.
Connect the unit to a mains socket of the same size as the plug used. Do not use adapters
between plug and socket.
3.
Ensure that the mains socket is equipped with a grounding line and safety devices.
4.
For continued protection against risk of fire, replace fuses only with components of the same type
and ratings.
5.
Electrical shock hazard. No user serviceable parts are inside. For servicing refer to qualified
personnel who are authorized only by the manufacturer.
6.
Never connect the plug to the mains socket with wet or damp hands. If liquids are poured onto
the unit during use, turn off the power switch and remove the plug immediately, before drying
the external panels.
7.
Always remove the plug from the socket by gripping it with the hand. Do not pull on the power cord.
8.
Ensure that the plug is secured to the machine by means of the metallic grid (see Figure 5-12) to
prevent accidental unplugging.
9.
To prevent risk of electrical shock, do not use alternate power plugs or adapters that disconnect
the green/yellow wire safety ground.
10. The electrical installation must be in compliance with electrical safety regulations and according
to voltage and current ratings specified on the labels on the back of the unit.
11. The X
TRA
system must be always grounded via the hospital electrical power source according to
the regulations in force in the country of use.
12. Electrical safety features are contained in
Appendix A: Safety Standards EN 60601-1-2
13. The use of accessories and cables other than those provided by LivaNova may result in increased
emissions or decreased immunity of the X
TRA
and X
VAC
devices.
14. Accessories and cables provided by LivaNova may only be used with the X
TRA
and X
VAC
devices.
The use with other equipment or system may result in increased emission or decreased immunity
of that equipment or system.
15. To prevent risk of electrical shock, shut OFF power and unplug the system from the electrical
outlet before performing cleaning procedures.
16. OPERATOR must not touch the relevant part and the PATIENT simultaneously:
– accessible contacts of connectors;
– contacts of fuse holders that are accessible during replacement of the fuse;
– parts inside an ACCESS COVER that can be opened without the use of a TOOL, or where a
TOOL is needed but the instructions for use instruct any OPERATOR other than SERVICE
PERSONNEL to open the relevant ACCESS COVER.
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...