Vacuum Module
X
TRA
Operator’s Manual
20077/021 US
13-9
•
Timeout communication failure
: Communication on the X
TRA
-X
VAC
RS-422 serial connection
has timed out (for example, because of a damaged cable or loose connection).
•
Serial communication failure
: A double communication error on the same communication
window on the X
TRA
-X
VAC
serial line has occurred.
When an alarm occurs (F1 to F11 and the two listed above) during or immediately after the prelink
phase, the X
TRA
user interface displays the message “Vacuum pump out of order”. The X
TRA
system
will continue working as if the X
VAC
module were absent.
When an alarm occurs during the T1 test (alarm F10 or F11) when the boot is over, the X
TRA
system
will consider the X
VAC
module absent without displaying any message.
When one of the two serial connection failures (listed above) occurs, the X
TRA
system switches from
remote control to local control. In addition, the red alarm LED lights up. Unlike with the other alarms,
in the case of the connection failures, the module continues to work (in local control).
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...