Installing the Disposables
X
TRA
Operator’s Manual
20077/021 US
4-3
Step-by-step Procedures for Installing Disposables
Step 1: Installing the Blood Collection Reservoir
The setup of the Blood Collection Reservoir can be performed by inserting it into its holder on the rear
of the unit or by using the separate optional holder, to be fixed to an IV pole in the proximity of the
surgical field.
1.
Swing open the reservoir holder and raise it to the desired height (see Figure 4-1).
Figure 4-1 Raising the Reservoir Holder
2.
Remove the Blood Collection Reservoir from its package.
3.
Keep the locking clip on the reservoir holder open, slide the edge of the reservoir lid into the slot
of the holder and push the reservoir into position. Release the locking clip (see Figure 4-2).
CAUTION
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...