Preoperative Sequestration (PPP and PRP)
X
TRA
Operator’s Manual
20077/021 US
10-15
Figure 10-10 Empty Phase of the Processing Screen
For a detailed explanation of the
Empty phase screen, read
“Empty Screen” on page 5-41.
The volume of the blood transferred from the bowl is displayed on the processing displet and
constantly updated by the system’s software. The empty speed can be adjusted from the processing
displet (from 25 ml/min to 1,000 ml/min).
When the bowl empties, the cycle has completed, the pump stops, and the Ready Screen is displayed.
Note:
If, for any reason, you need to stop a cycle before it has completed, please refer to “Stop
After the air sensor detects that the bowl has been emptied and the Empty cycle has been completed,
you may repeat the procedure as directed or terminate sequestration processing and convert the
system for intraoperative blood recovery (see “Finishing Preoperative Separation and Preparing for
Blood Recovery” on page 10-19).
8.
To begin another processing cycle, maybe using a different protocol, repeat the process from
Step 1 on page 10-11. Otherwise, to end the current case, touch the “End Case” button.
Ending the current case will display the End of Case Screen. Follow the instructions in the section
“Ending the Current Case” on page 5-14 to start a new case, or unload the pump loop and shut down
the system.
CAUTION
Air must be collected and be available to be returned to the centrifuge bowl
during the Empty cycle. Otherwise, plasma may be accidentally drawn back
into the centrifuge bowl. Ensure that the bag with the air remains attached to
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...