Installing the Disposables
4-10
20077/021 US
X
TRA
Operator’s Manual
Step 5: Connecting Wash Lines to the Saline solution Containers
1.
Remove the RBC bag from the tray and hang it on the upper left hanger of the IV pole.
2.
Connect the wash lines (yellow) to the wash solution (see Figure 4-11).
a. Close the slide clamp of one of the two yellow wash lines (when using only one bag of saline
solution)
b. Hang the wash solution bags on the lower hangers of the IV pole.
c. Using aseptic technique, spike the wash line into a bag of saline solution
Figure 4-11 Connecting the Wash Lines (Yellow) to the Washing Solution
Note:
Please refer to
Appendix D: Recommended Fluid Bag Configurations
for recommended
configurations for hanging the maximum number of fluid bags on the X
TRA
IV poles. These
configurations ensure that the X
TRA
remains stable under normal operating conditions.
Step 6: Hanging the Waste Bag
1.
Remove the waste bag from the tray.
2.
Hang the waste bag on the two lower hangers located on the right panel of the machine. If
Plasma Sequestration is to be performed, skip next step.
CAUTION
Use aseptic technique when installing disposables.
CAUTION
Sterile 0.9% normal saline, USP (injectable or approved for cell processing) is
typically used as a wash solution. Other solutions intended for intravenous
use that have been approved by the FDA and have documentation available
to show the component is safe may be used.
i
American Association of Blood Banks.
Standards for Perioperative Autologous Blood
Collection and Administration
. 3rd Edition. Bethesda, MD. 2007 Reference Standard
5.4.4 (Addition of Drugs and Solutions)
CAUTION
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...