Preoperative Sequestration (PPP and PRP)
X
TRA
Operator’s Manual
20077/021 US
10-19
Finishing Preoperative Separation and Preparing for Blood Recovery
To finish preoperative sequestration and prepare for blood recovery, do the following:
1.
Remove any air from the PRP bags.
2.
Clamp both PRP bags. Detach the PRP bags from the plasma sequestration lines.
3.
using the caps and plugs provided, cap and plug the PRP bags and lines. This will decrease the
chance of contamination and any bio-hazard caused by dripping.
4.
Reinfuse autologous plasma according to hospital guidelines and/or AABB standards.
5.
Unclamp the inlet tube to the waste bag.
6.
Clamp the blood inlet line from the whole blood bag to isolate the patient line and unclamp the
line from the reservoir.
7.
Connect the Wash line to the wash solution.
a. Hang the wash solution bags on the lower hooks of the IV pole.
b. Using aseptic technique, spike each Wash line into a bag of wash solution. Unclamp at least
one line.
8.
If you are using a new reinfusion bag:
a. Remove the cap from the reinfusion bag.
b. Remove the cap from the tubing line and twist connect to the reinfusion bag
c. Hang the new reinfusion bag.
WARNING
To reduce risk of air embolism, remove all air from the plasma bags
before handing the bag over for reinfusion.
WARNING
WARNING
i
American Association of Blood Banks.
Standards for Perioperative Autologous Blood
Collection and Administration
. 3rd Edition. Bethesda, MD. 2007 Reference Standard
5.1.8 (Handling, Storage, Transportation)
WARNING
Do not reinfuse the PRP back to the patient if the X
CAUTION
Make sure that the waste line is unclamped before starting blood salvage
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...