Processing
X
TRA
Operator’s Manual
20077/021 US
5-35
Set the Active Protocol
By default, when the X
TRA
is powered on, the active protocol is the protocol which has been
designated as the wakeup protocol. To change the wakeup protocol, refer to “Changing the Wakeup
Protocol” on page 9-6.
If you want to use a protocol for the next cycle that is different from the currently active protocol, do the
following:
1.
Touch the Menu button from the Ready Screen to enter the Menu Screen.
2.
From the left side of the Menu Screen, touch the Protocol/Mode tab.
3.
Using the protocol drop-down menu at the top of the Protocol/Mode tab, select the desired
protocol:
a.
Touch the triangle menu button located on the right side of the protocol drop-down
menu. A list of available protocols will be displayed.
b. Touch the name of the desired protocol to close the menu and set it as the active protocol.
(Touching the triangle menu button a second time without selecting a protocol from the list will
close the Protocol menu without changing the active protocol.)
4.
Touch the Save Modifications button to make all changes in the Protocol/Mode tab take
effect. If the screen is exited by touching the Close button, or is interrupted by a warning or
alarm, then the modifications are not saved.
Set the Active Operating Mode
Do the following to set the active operating mode:
1.
Touch the Menu button from the Ready Screen or any Processing Screen to enter the Menu
Screen.
2.
From the left side of the Menu Screen, touch the Protocol/Mode tab.
3.
Using the mode drop-down menu at the top right of the Protocol/Mode tab, select the desired
operating mode:
a.
Touch the triangle menu button located on the right side of the Mode menu. A list of
available modes for the selected protocol will be displayed.
b. Touch the name of the desired mode to close the menu and set it as the active operating
mode. (Touching the triangle menu button a second time without selecting a mode from the
list will close the Mode menu without changing the active mode.)
4.
Optionally, enable or disable any automations desired for the selected mode using the controls in
the area immediately below the mode drop-down menu. Refer to
for a full description of each.
Touching the button for the Autostart and Continue automations will toggle its
state. When the automation is ON (active), its button will appear inset and green;
when the automation is OFF, its button will return to the normal raised
appearance. When the Autostart and/or Continue automations are enabled, its icon will appear in
the status area of every screen.
When Manual mode is selected, the RBC detector and Wash-Empty automation may be enabled
or disabled by touching the appropriate check box. When the box is checked, the detector/
automation is ON.
5.
Touch the Save Modifications button to make all changes in the Protocol/Mode tab take
effect. If the screen is exited by touching the Close button, or is interrupted by a warning or
alarm, then the modifications are not saved.
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...