Vacuum Module
13-6
20077/021 US
X
TRA
Operator’s Manual
Use of the X
VAC
System Through the X
VAC
Control Panel (Local Control)
The X
VAC
vacuum module can be controlled using its built-in control panel only when the touch screen
user interface of X
TRA
is not available, in the following two cases:
1.
When the X
TRA
system is booting up
2.
When a fatal error occurred on the X
TRA
system
X
VAC
Control Panel
The control panel (see Figure 13-6) is located on the front of the X
VAC
module and is used to control
the vacuum when the X
TRA
user interface is not available (local control) (i.e., when the X
TRA
system is
booting up or when a fatal error occurred on the X
TRA
system).
Figure 13-6 X
VAC
Control Panel
When operating in local control, the display of the X
VAC
control panel will show the “not linked” icon:
Vacuum Pump On/Off Button
Pressing the vacuum on/off button will toggle the vacuum pump on and off when it is operating
in local control. The current state of the vacuum pump is indicated by the green vacuum status
LED (light=ON; dark=OFF).
Vacuum Level Setpoint Selector (Up/Down buttons)
Pressing the increase (+) and decrease (-) buttons will change the vacuum level setpoint
(mmHg) for the selected operating mode.
Mode Buttons
Pressing the INTRA and POST mode buttons changes the X
VAC
’s operating mode between
intraoperative and postoperative, respectively. A green LED next to each button indicates
the current mode (LED on or blinking next to a button indicates that the module is working
or is ready to work in that mode).
Mute Button
Pressing the mute button silences the system beep when an alarm occurs.
Alarm Indicator LED
The red LED, located near the yellow warning icon at the upper-right corner of the control
panel, turns on to indicate that an alarm or error has occurred.
WARNING
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...