Vacuum Module
13-8
20077/021 US
X
TRA
Operator’s Manual
Alarms
The section lists the possible alarms, and their causes and effects, that may occur during the operation
of the X
VAC
module.
List of Fatal Alarms
Alarm Management in Local Control
When one of the alarms listed in Table 13-1 occurs while the X
VAC
module is operating in local control
(i.e., when the X
TRA
system is booting up or when a fatal error occurred on the X
TRA
system), the
following actions are taken:
• Vacuum pumps are switched off
• The red alarm LED is turned on
• The system beep is sounded
• The code corresponding to the alarm is shown on the X
VAC
LCD display
• The system cuts off the supply of the 230VAC Intra pump by the activation of the safety relay
(only for alarms F1, F2, F3, F4, F5)
To stop an alarm, the operator must power off the module and resolve the cause of the alarm.
If the EEPROM failure alarm (F11) occurs due to the failure of calibration coefficients test, the system
automatically repairs the problem by overwriting the wrong coefficients with a default set of
coefficients.
Alarm Management in Remote Control
For the X
VAC
module, every point made in “Alarm Management in Local Control” on page 13-8 is also
true when operating in remote control.
For the X
TRA
system, there are two other causes of alarms when operating in remote control:
Error Service Code(s)
Error Family
Operator Response
F1
Over temperature (in the inner module)
• Switch the system off
• Ensure that the fan in the rear of the
module is not occluded
• Switch the system on
• If the problem persists, call the
authorized technical service
F2
F3
Overpressure
• Switch the system off
• Switch the system on
• Call the authorized technical service
F4
F5
Error in pressure measurement
• Switch the system off
• Switch the system on
• Call the authorized technical service
F10
F11
Hardware Failure
• Call the authorized technical service
Table 13-1 List of Error Service Codes
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...