Configuring X
TRA
8-4
20077/021 US
X
TRA
Operator’s Manual
Configuration Mode Screen Tabs
Setting the Enabled Features
The first tab (from the top) of the Configuration Mode Screen is the Features tab (see Figure 8-3).
The optional features of the X
TRA
system may be enabled and disabled from this tab.
Figure 8-3 Features Tab of the Configuration Mode
As mentioned under “Predefined Feature Sets” on page 8-3, from this tab it is possible to select a
predefined configuration set using the Feature Set button at the top of the screen.
Indicators
The following three settings relate to the wash quality indicators as described in
.
Hct indicator.
The Hct indicator option turns the Hct (hematocrit) indicator ON (enabled) and
OFF (disabled). It must be enabled for features related to the Hct indicator to be available during
processing (such as for the “Hct in” and “Hct out” figures to appear on processing screens and be
stored with the case data). If the Hct indicator is disabled, the Hct indicator will NOT be
calibrated during setup.
Supernatant removal indicator.
This option turns the supernatant removal indicator ON
(enabled) and OFF (disabled). It must be enabled in order to have the possibility to activate the
Waste line color indicator.
Waste line color indicator.
This option turns the three-color waste line indicator ON (enabled)
and OFF (disabled).
Protocol Set
Popt
enabled
enabled
Pstd
disabled
enabled
Pfat
enabled
enabled
Post-op
disabled
disabled
PPP
disabled
disabled
PRP1
disabled
disabled
PRP2
disabled
disabled
Emergency key
enabled
enabled
No Wash option
disabled
disabled
Rapid Transfer option
disabled
disabled
User Protocols
disabled
disabled
Wake up
ATS (Intraoperative) Mode of
Operation
1 touch
1 touch
Tab
Feature/Function
Basic
Advanced
Table 8-1 Comparison of the Predefined Feature Sets (Continued)
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...