System Description
3-22
20077/021 US
X
TRA
Operator’s Manual
The pump loop cartridge includes the pump loop which is placed over the process pump head, and it
will self-load when the pump starts to rotate.
The three color-coded tubes extend from a 3-way connector attached to the pump loop cartridge and
are mounted on a tubing organizer which assures the correct tubing is routed through the correct
clamp. The tubing from the pump to the centrifuge bowl passes through the air sensor.
The pump tubing is connected to the inlet port on the centrifuge bowl. The tubing from the centrifuge
bowl outlet port connects to the 10-liter Waste bag (which is not pre-connected).
Rapid Latham Centrifuge Bowl
The Rapid Latham centrifuge bowl (see Figure 3-32) is designed specifically for concentration and
separation of blood components and for high removal of waste fluids collected during the intra- and
postoperative procedure.
Figure 3-32 Cross-Section of Latham Centrifuge Bowl
The conic-sided design of the Rapid Latham bowl, together with the double RBC detector system,
allows a superior separation of the blood components and the rapid processing of a high quality red
blood cell product.
The patented bowl clutch mechanism allows the bowl to be installed on the centrifuge plate with an
easy push action.
The centrifuge bowl is available in four sizes: 55 ml, 125 ml, 175 ml, and 225 ml. All bowls have the
same external dimensions. The size of the bowl is signified by the color of the ring on the bowl collar:
The large variety of bowl sizes available allows a dedicated approach to the problems of blood
recovery, providing a specific intervention in all surgical specialties and situations.
Color
Fluid Carried
Clamp Name
Blue
Fluid collected in reservoir
Fill clamp
Yellow
Wash solution
Wash clamp
Red
Washed RBCs
Empty clamp
Table 3-1 Color-Coded Tubing and Clamp Association
Bowl Size
Color
55 ml
White
125 ml
Green
175 ml
Red
225 ml
Blue
Table 3-2 Different Bowl Size Colors
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...