Introduction and Safety
1-14
20077/021 US
X
TRA
Operator’s Manual
67. Although it is possible to configure some of the acoustic notifications that the machine makes,
any modification on the audible signals might make the operator take longer to realize the
machine has made a notification.
68. It is possible to temporarily deactivate some of the controls that the machine does. The operator
can make use of this possibility, under his own responsibility, therefore directly managing those
controls.
69. At the end of installation verify that the bowl outlet is not crushed or obstructed.
70. In the case of malfunction of the integrated vacuum pump, use an alternative vacuum source,
i.e., the vacuum present in the operating room, possibly connected to a pressure regulator.
71. If the values regarding volumes, hematocrit, supernatant removal and waste color line are
important for the patient, it will always be necessary to use other standard measurement
instruments of the hospital.
Operating Conditions
Warning
1.
Do not use the unit in the presence of inflammable anaesthetic gases. do not expose to heat
sources. Avoid any operating situation where the blood may be exposed to temperatures
exceeding 37 °C (98.6 °F).
Cautions
1.
When the unit is operating, the following conditions must be met:
- Room temperature: 10 - 35 °C (50 - 95 °F)
- Relative humidity: 30-75%
- Atmospheric pressure: 80 - 106 kPa
2.
Conditions that cause the unit to overheat when operating must be avoided.
Adverse Effects
The company is not aware of documented adverse effects when this device is used according to this
Operator's Manual.
Installation of the Unit
Installation
The X
TRA
and X
VAC
systems are medical electrical equipment and need special precautions regarding
electromagnetic compatibility (EMC). They need to be installed and put into service according to the
EMC information provided in
Appendix A: Safety Standards EN 60601-1-2
.
Portable and mobile radio frequency (RF) communications equipment can affect the X
TRA
and X
VAC
.
WARNING
should not be used adjacent to or stacked with
other equipment. However, if adjacent or stacked use is necessary,
the X
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...