
Processing
5-50
20077/021 US
X
TRA
Operator’s Manual
End of Case Screen
The End of Case Screen (see Figure 5-48) appears after touching the “End Case” button from the
Ready Screen. From the End of Case Screen, the operator may empty the RBC line and unload the
pump loop, as well as review, export, and print the case tally data for present and past cases.
The Vacuum displet is the only displet present in the End of Case Screen.
For instructions on ending the current case and unloading the disposable, refer to the section “Ending
the Current Case” on page 5-14.
Figure 5-48 End of Case Screen
The information and interface buttons related to viewing and exporting case data from the End of
Case Screen also appear in the Tally Tab of the Menu Screen. For a detailed description, read “Tally
Tab” on page 5-44.
For instructions on exporting and printing case data, refer to
Chapter 11: Data Download Option
1. Empty Button
The Empty button may be used to empty the RBCs present in the empty line without
unloading the pump loop tubing. Touching the Empty button displays the message “RBC line
emptying....” at the top of the screen while the RBC line is being emptied.
2. Unload Button
Touching the Unload button ends the current case and begins the auto-unload process. Once
the pump loop is unloaded, the “Start New Case” button appears. After the Unload button is
touched, it is impossible to return to the Ready screen without beginning a new case, and it
is impossible to continue the current case at the next system power-on (the Retain button will not be
available). Refer to section “Ending the Current Case” on page 5-14 for complete instructions.
3. Close Button
Touching the Close button closes the End of Case Screen without ending the current case
and returns to the Ready Screen where further processing cycles may be initiated.
“Start New Case” Button
The “Start New Case” button appears after the case has ended and the pump has been
unloaded. Touching the “Start New Case” button displays the Setup Screen and begins a
new processing case.
1
3
1 Empty RBC button
3 Data destination menu
2 Unload button
4 Close button
2
4
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...