A-2
20077/021 US
X
TRA
Operator’s Manual
Guidance and manufacturer's declaration – electromagnetic immunity
The X
TRA
is intended for use in the professional healthcare facility environment specified below. In order to prevent
adverse advents to the patient and operator due to electromagnetic disturbances, the X
TRA
System must not be operated
outside its intended EMC environment. Furthermore, the X
TRA
must not be operated if the enclosure, cables or measures
for electromagnetic shielding are damaged.
Phenomenon
Basic EMC standard or
test method
IMMUNITY TEST LEVELS
Professional healthcare facility environment
ELECTROSTATIC DISCHARGE
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
Radiated RF EM fields
IEC 61000-4-3
3 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
Proximity fields from RF wireless
communications equipment
IEC 61000-4-3
See Table A‐3
RATED power frequency magnetic
fields
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
Table A-2 Enclosure port
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...