Preoperative Sequestration (PPP and PRP)
10-24
20077/021 US
X
TRA
Operator’s Manual
Empty Phase
The Empty phase of the PRP2 protocol, during which the remaining blood components in the bowl are
transferred to the RBC bag, is identical to that of the PPP and PRP1 protocols detailed in “PPP” on
page 10-22. Before the Empty phase begins, the operator will be prompted to close the PRP line and
open the PPP line so air can enter the bowl during emptying.
Protocol Parameters (PPP and PRP)
The table below contains the parameters recommended by LivaNova for the PPP and PRP protocols,
according to the bowl size used.
Note:
Every time a new case is started, factory procedure parameters are reinitialized to their
default values, even if the unit is not shut down and restarted.
Protocol
Parameters
Range
Step
Unit
Factory values
X/55
X/125
X/175
X/225
PPP
Fill Flow
10 - 100
10
ml/min
20
50
50
60
Final Fill Flow
i
i
These parameters can be modified run time only and modifications will be lost at every new cycle.
10 - 100
10
ml/min
20
50
50
60
Fill Centrifuge Speed
ii
ii
These parameters, Centrifuge Speed in the Fill and in Spill phases, are related to each other: Spill Centrifuge Speed must always be lower
than Fill Centrifuge Speed.
1500 - 5600 100
rpm
5600
5600
5600
5600
Empty Flow
25 - 1000
25
ml/min
100
100
100
100
PRP1
Fill Flow
10 - 100
10
ml/min
20
50
50
60
Final Fill Flow
10 - 100
10
ml/min
20
50
50
60
Fill Centrifuge Speed
2400 ˜5600 100
rpm
5600
5600
5600
5600
Spill PPP Flow
10 - 100
10
ml/min
10
10
10
40
Spill PPP Vol
0 - 400
1
ml
27
30
35
100
Spill PRP Flow
10 - 100
10
ml/min
10
10
10
20
Spill PRP Vol
0 - 400
1
ml
25
35
35
55
Spill Centrifuge Speed
2400 - 5600 100
rpm
2400
2400
2400
2400
Empty Flow
25 - 1000
25
ml/min
100
100
100
100
PRP2
Fill Flow
10 - 100
10
ml/min
20
50
50
60
Final Fill Flow
10 - 100
10
ml/min
10
30
30
60
Fill Centrifuge Speed
2400 - 5600 100
rpm
5600
5600
5600
5600
Spill PPP Flow
10 - 100
10
ml/min
10
10
10
40
Spill PPP Vol
0 - 400
1
ml
27
30
35
100
Spill PRP Flow
10 - 100
10
ml/min
10
10
10
20
Spill PRP Vol
0 - 400
1
ml
20
25
25
45
Spill Centrifuge Speed
2400 - 5600 100
rpm
2400
2400
2400
2400
Empty Flow
25 - 1000
25
ml/min
100
100
100
100
Table 10-2 PPP and PRP Protocol Parameter Summary
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...