Special Cycles
X
TRA
Operator’s Manual
20077/021 US
6-5
the saline solution from the bowl towards the RBC bag. The centrifuge does not spin during this
operation.
The Prime IV Screen is displayed throughout the cycle (see Figure 6-4).
Figure 6-4 Prime IV Screen (Paused)
The Prime IV Screen tracks the volume (ml) of saline solution which has been pumped from the bowl
towards the RBC bag.
The Prime IV Screen is initially paused. The operator must touch the Play button to
initiate the cycle. Touching the Pause button will stop the pump rotor. To resume the
Prime IV cycle while paused, touch the Play button.
Touching the Close button will stop the Prime IV cycle and return to the Ready Screen. The
Prime IV cycle may later be resumed from the Ready Screen at the point at which it was
interrupted, before starting blood processing.
The cycle ends automatically once the whole amount of the saline solution has been pumped into the
RBC bag and air is detected in the line by the air sensor, at which point the Ready Screen is displayed.
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...