Special Cycles
6-8
20077/021 US
X
TRA
Operator’s Manual
Connecting the Blood Administration Set to the X
TRA
RBC Bags
If it is necessary to establish a continuous loop for reinfusion of blood processed by the X
TRA
, follow
these steps:
1.
Using aseptic technique, spike one of the unused ports of the X
TRA
RBC bag (either of the two
unused ports may be used) with either:
◊
a blood administration set which has an integral microaggregate (20-40 micron) transfusion
filter (see Figure 6-6 Top).
OR
◊
a separate microaggregate (20-40 micron) transfusion filter (see Figure 6-6 Bottom).
2.
Make certain that the primary clamp is in the open position. Prime and de-bubble the blood
administration and microaggregate filter set according to manufacturer's recommendations. Then
close the distal roller clamp on the blood administration line.
3.
Clamp off a secondary RBC bag.
4.
Attach the second leg of the “Y”-type blood administration set to one of the remaining ports of
the secondary RBC bag.
5.
Close the primary clamp. This clamp should remain closed during processing with the X
TRA
system.
6.
After a unit of packed red blood cells has been emptied into the X
TRA
RBC bag, open the primary
clamp to allow the cells to flow into the secondary RBC bag. Then close the primary clamp prior
to initiating reinfusion to the patient.
Note:
The primary clamp on the leg of the “Y”-type blood administration set attached to the RBC
bag should remain closed during processing and/or reinfusion.
7.
The distal end of the blood administration set is now ready to be connected to a suitable venous
catheter by the anesthesiologist.
WARNING
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...