Installing the Disposables
X
TRA
Operator’s Manual
20077/021 US
4-9
Figure 4-8 Seating the Pump Loop Cartridge (Left); Latching the Clamp Lid (Right)
Note:
Do not force the pump loop down the rotor since it will load automatically when the LOAD
PUMP button is pressed.
13. Make sure that the tubing is correctly seated into the air sensor (see “C” in Figure 4-6).
14. Close and latch the clamp lid (see the right image of Figure 4-8).
15. Make sure the waste line is properly seated into the waste fluid transparency indicator.
16. Close the centrifuge lid.
17. Connect the fill line (blue) to the outlet port of the reservoir covered with a blue cap . . .
a. located on the reservoir lid, marked WASH, for a Blood Collection Reservoir TOP (see
b. located on the bottom of the reservoir for a Blood Collection Reservoir BOTTOM. In this case
once connected, open the slide clamp (see Figure 4-10).
Figure 4-9 Connecting the Bowl Set to the Top of the Blood Collection Reservoir
Figure 4-10 Connecting the Bowl Set to the Bottom of the Blood Collection Reservoir
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...