Processing
X
TRA
Operator’s Manual
20077/021 US
5-47
Figure 5-45 Protocol/Mode Tab of the Menu Screen
The Protocol/Mode tab is divided into two areas: the protocol area on the left which contains the
protocol selection menu, parameters and automations related to the selected protocol, and the mode
area on the right which usually contains the operating mode selection menu and automations related
to the selected mode.
1. Protocol Area
The protocol area contains the controls used to select and modify the active processing protocol. At
the top is the protocol drop-down menu, which is used to select the active protocol. Below that appear
parameters and controls specific to the selected protocol. If the User Protocols option is enabled,
buttons which can be used to create, rename, and delete custom protocols, as well as designate the
wakeup protocol are displayed at the bottom of the screen. Modifying the parameters of protocols and
managing custom protocols are described in
Chapter 9: Programmability Option
, specifically “The
Protocol/Mode Tab” on page 9-7.
The active protocol cannot be changed during a processing cycle.
2. Mode Area
When the Protocol/Mode tab is entered after the pump has been loaded, the mode area contains the
drop-down operating mode selection menu used to select the active operating mode.
Controls for any automated functions (such as Autostart, Continue, RBC Detector, and the
Wash-Empty functions) which are available for the selected mode will appear directly beneath the
drop-down Mode menu. For details on each automation, refer to
Chapter 7: Automated Functions
Touching the button for the Autostart and Continue automations will toggle its state.
When the automation is on, its button will appear inset and green; when the automation
is OFF, its button will return to the normal raised appearance. When the Autostart and/
or Continue automations are enabled, its icon will appear in the status area of every screen.
When Manual mode is selected, the RBC detector and Wash-Empty automation may be enabled or
disabled by touching the appropriate check box. When the box is checked, the detector/automation is
ON.
When the Protocol/Mode tab is entered from the Setup Screen (before the pump has been
loaded), the mode area contains a single “Bowl” button. Each time the button is touched, it
toggles the bowl size for the processing case (55 ml, 125 ml, 175 ml, or 225 ml). This
allows the operator to:
• view the protocols available for each bowl size (also factory protocols)
• create new user protocols and set their parameters for each bowl size.
• to change the wake up protocol for each bowl size.
1
2
1 Protocol area
2 Mode area
3 Save Modifications button
3
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...