Troubleshooting
X
TRA
Operator’s Manual
20077/021 US
14-19
RBC bag empty.
Bowl not filled
Warning
The volume inside the RBC bag
is not sufficient to properly fill
the bowl.
• Make sure the line is
connected and not clamped,
then press PLAY to resume.
• Otherwise, if the bowl is
sufficiently packed with RBCs,
you may decide to press
WASH to proceed to the
Wash phase or 2xWASH to
use twice the saline solution
volume.
WARNING: Inadequate filling of
the bowl may lead to lower
than expected hematocrit and
removal of waste components.
If the problem persists, call the
authorized technical service.
RBC bag empty
RBC bag empty.
Bowl not filled (SEQ)
Warning
The sensor is detecting air in
line.
• The empty line might be
clamped or occluded. Remove
the occlusion and press PLAY
to resume.
• The RBC bag might be empty.
The spill phase cannot be
started if the bowl has not
been properly filled with
RBCs. If you intend to
proceed, connect a new
whole blood bag and press
PLAY to resume.
• If you intend to end the
sequestration, press EMPTY
to recover the bowl content
into the RBC bag.
RBC bag empty
RBC bag full. Empty
or replace it.
Alarm
The volume transferred to the
RBC bag exceeds the pre-set
safety level.
Empty or replace the RBC bag,
then press PLAY to resume. The
bag volume counter will be set
to Zero.
WARNING: Not following these
instructions may expose the
operator to biohazard risk and
compromise the current case.
RBC bag full
Message Text
Warning or
Alarm
Buttons Available in the
Button Bar During
Warning/Alarm
Operator Response
Data Management
Text
Table 14-2 List of Warnings and Alarms Which Might Occur During Processing (Continued)
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...